CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System Recall
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
FDA recall Class II
Ongoing
FDA lists this Class II recall for CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-3334-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Aug 18, 2026
- FDA report date
- Oct 7, 2026
- Quantity in commerce
- 34
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System
Affected model & UDI checker
Search the exact identifiers published for Z-3334-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 40122716 | CHARLOTTE CLAW Screw, REF: 40122716, intended for use with CLAW II Polyaxial Compression Plating System | Check source notice | Not specified |
FDA code information
UDI-DI: 00840420109705, Lot: 1826666, Distributed: 18-Aug-2025 through 09-Apr-2026
Why was this product recalled?
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
What actions did WRIGHT MEDICAL TECHNOLOGY, INC. take?
On 8/13/2026, recall notices were mailed and emailed to customers who were asked to do the following: 1) Review the product issue and risks as communicated in this Recall Notice and communicate the issue as appropriate in your organization. 2) Sign and return the enclosed business reply form by email to [email protected] 3) Upon receipt of the completed business reply form, firm will arrange for product return. 4) If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. Inform firm if any of the subject devices have been distributed to other organizations, including contact details, so that firm can inform the recipients appropriately. b. If you are a distributor, notify your affected customers. For questions or concerns, contact the firm at [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA classification recorded
FDA notice posted
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-3334-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 19, 2026
- openFDA device recall record · Source release Nov 14, 2026 · Retrieved Nov 19, 2026