Elevation Spine appears in FDA premarket records as the applicant for the SABER-C System, an intervertebral body fusion device in Class II under 21 CFR 888.3080, product codes OVE, ODP and KWQ. The supplied collection holds one linked 510(k) record, K263049, decided September 23, 2026 with decision code SESE, meaning substantially equivalent. That single record is a snapshot of the supplied filings, not a measure of the company's total portfolio or commercial activity. [1][2]

The SABER-C System as documented

The company's own site describes SABER-C as a cervical fusion system made of a zero-profile anterior cervical plate, an interbody and fixation, with plate and titanium interbody sizes listed as 12x15mm and 14x17mm at 5-9mm height and 6 degrees of lordosis. The same publisher states that when Saber-C is used with spikes, supplemental fixation is required. A separate page on the same site lists SABER-C alongside the AVIA ACDF Fixation System and the SABER-XA Expandable ALIF System. [4][5]

Filing history in FDA records

K263049 was received August 31, 2026 and decided September 23, 2026 as substantially equivalent, with a summary statement and a special clearance type. The FDA letter for that submission lists product codes OVE, ODP and KWQ and routes correspondence through a regulatory consultant rather than a company address. [1][2]

Earlier clearances for the same device name appear in the numbered sources. K260660 was dated February 27, 2026 and received March 2, 2026, with a decision letter dated April 27, 2026. K250540 was dated and received August 1, 2025, with a letter dated August 8, 2025. K190885, the earliest of these, was dated July 12, 2019 and received July 15, 2019. [6][8][9]

What the submissions say about testing

The K263049 letter compares the subject device with predicate K260660 and describes the testing carried over from that filing. Its bench testing discussion covers compression, compression shear and torsion under ASTM F2077, along with expulsion and subsidence under ASTM F2267. The original letter gives the individual test conditions and the scope of the submission’s indications change. [2]

The K250540 letter describes bench testing of the Saber-C System covering axial screw pullout and insertion torque under ASTM F543 and plate pushoff. It compares materials, manufacturing, sterilization and packaging with the primary predicate Saber-C (K190885). The K190885 letter describes mechanical testing under ASTM F2077-14 and ASTM F2267-11, expulsion testing and a cadaver implantation study of the plate and spacer assembly with spiked fixation. [6][8]

Applicant identity and limits of the record

The applicant of record is Elevation Spine, identified in the K263049 letter as Elevation Spine, Inc. Correspondence addresses in the filings are those of regulatory consultants, and the K263049 database entry lists a Monterey, California address; neither is established here as a headquarters. No parent, brand owner or corporate structure is verified in the public documents, and no recall records are present. [1][2]