FDA enforcement records identify Olympus Corporation of the Americas as the recalling firm behind several 2026 Class II device recalls covering reusable optical telescopes, electrosurgical hemostasis probes and a video accessory cable. The filings describe discrete model and catalog numbers, lot or serial scope, unit quantities and distribution patterns, and each carries a stated reason for recall. The records below are a snapshot of those filings, not a measure of the company's total portfolio or safety performance. [1][2][3][4][5]

Telescope sterilization-parameter recalls

Two related filings cover the Olympus M3 Gold Autoclavable Lateral Telescope 70, model M3-70A, and the M3 Gold Autoclavable Foroblique Telescope 30, model M3-30A. Both cite identified discrepancies between the sterilization parameters listed in the Instructions for Use and those listed in the Japanese package insert. The lateral telescope filing covers 50 units and all serial numbers; the foroblique filing covers 72 units and all serial numbers. Both were initiated August 24, 2026, classified Class II, and distributed internationally to Japan. [1][2]

FDA recall database pages separately list each telescope as a Class 2 device recall, and an event-level search returns two related medical device recalls under event 99779. The records do not state a root cause beyond the parameter discrepancy or describe any reported patient harm. [6][7][8]

BiCOAG hemostasis probe filings

The BiCOAG Hemostasis Probe 7Fr Coaxial Connector, catalog CD-B620LA, and the 10Fr version, catalog CD-B610LA, were recalled over a potential for detachment of the ceramic tip due to manufacturing without the designated amount of adhesive. The filings cover all unexpired lots, with 2,142 units of the 7Fr probe and 404 units of the 10Fr probe. Both were initiated August 19, 2026, classified Class II on September 18, 2026, and distributed to the US, Canada, Latin America, Singapore, Hong Kong and the EU. [4][5]

OLYSENSE Hub accessory cable

A separate filing covers the Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, model P580-025, material number OS-DPDPC25-G, described as an accessory used with the OLYSENSE Hub System. The stated reason is that a video dropout or blackout may occur in the high-power operating mode when using that particular accessory cable. The recall covers 543 cables, was initiated August 27, 2026, classified Class II on September 29, 2026, and was distributed in the US and Poland. [3]

Corporate identity and published materials

The recalling firm address in the FDA filings is 3500 Corporate Pkwy, Center Valley, PA, which is a filing address rather than a verified headquarters. FDA establishment registration and device listing records reference an Olympus establishment, and the company's own website describes Olympus as developing and delivering medical technology solutions and lists Medical Systems among its business areas. Those pages do not establish ownership, certifications or commercial availability of the recalled models. [1][9][10]